Limit of Quantification (LOQ) – What Is It?
The limit of quantification (LOQ) is the lowest concentration of an analyte that a given method can quantify with acceptable precision and accuracy. Above this value, the laboratory can report a specific numerical result; below it, it cannot. In the report, the result is then recorded as “< LOQ” along with the limit itself, for example, “< 0.01 mg/kg.”
It should not be confused with the limit of detection (LOD), which is lower and merely indicates the ability to detect the presence of the analyte. In the EU criteria for the performance of methods for trace elements in food, the LOD is defined as three-tenths of the LOQ, but this is not a rule that applies to all analyses. In older documentation, the abbreviation LOD was sometimes used with a different meaning; therefore, when comparing documents from different years, it is worth checking how the parameter was defined.
In B2B trade, the LOQ determines how to interpret an analysis certificate. A result of “not detected” without a specified limit is incomplete information—it is unclear whether the method was sufficiently sensitive relative to the limit applicable to the product.
Frequently Asked Questions
How is the limit of quantification determined?
There is no single universal procedure. In regulations concerning trace elements, the LOQ is defined as the lowest concentration that can be measured with reasonable statistical certainty; with constant accuracy and precision in the concentration range around the LOD, it corresponds to ten times the standard deviation of the mean from at least twenty determinations of a blank matrix sample.
Other methods include using the signal-to-noise ratio, calculating from a calibration curve, or adopting the lowest validated spiking level with acceptable recovery. Results from different procedures can vary several-fold; therefore, the LOQ value is meaningful only when it is known how it was determined. It is worth remembering that this is a conventional limit—in metrology, it is noted that results below this limit are not devoid of informational value, and a better criterion for suitability is often the measurement uncertainty as a function of concentration.
Why should the LOQ be lower than the product limit?
Because a method with an LOQ close to the limit does not allow for a determination of compliance. For lead in food, the required LOQ depends on the maximum permissible level (ML): for an ML of 0.01 mg/kg or less, an LOQ at the ML level is sufficient; for levels between 0.01 and 0.02 mg/kg, two-thirds of the ML is required; for levels between 0.02 and 0.1 mg/kg, two-fifths of the ML is required; and for levels of 0.1 mg/kg or higher, one-fifth of the ML is required. For cadmium, mercury, and inorganic arsenic, the criterion is two-fifths of the ML below 0.100 mg/kg and one-fifth above that value.
The method used to assess compliance should meet the criterion specified for the given limit. An LOQ that is too high renders the result unusable, even if the analysis was performed correctly.
What does a result of “< LOQ” mean on an analysis certificate?
It means that the concentration of the analyte is lower than the lowest value that the method can reliably measure. It does not mean zero—the analyte may be present below the method’s detection limit.
For total parameters—particularly dioxins and PCBs—how results below the LOQ are handled affects the final result. The approaches used are the lower limit (zero), upper limit (a value equal to the LOQ), and midpoint (half the LOQ). The same sample can therefore yield different totals, which is why it is advisable to specify the adopted convention in the specifications.
Is the LOQ the same as the reporting limit?
Not always, although they often coincide. The reporting limit (RL) is the lowest value that a laboratory reports—it may be equal to the LOQ or higher. In pesticide residue analysis, EU guidelines link it to the lowest validated level; a typical value in multi-component analysis is 0.01 mg/kg. When comparing results from two laboratories, this difference can sometimes be a source of misunderstanding.
How should LOQ requirements be specified in the specifications?
It is advisable to specify the maximum acceptable limit of quantification for each parameter or to refer to the criteria used in official controls. The statement “lead: max. 0.10 mg/kg, LOQ max. 0.02 mg/kg, ICP-MS method” is unambiguous, whereas “lead: not detected” says nothing about the sensitivity of the test. This is particularly important for raw materials sourced from multiple suppliers, where certificates come from laboratories with methods of incomparable sensitivity.